THZ & TDZ
Smart HART
®
Temperature Transmitters
*High-accuracy measurements are achieved by using a 4-wire, 1000 ohm platinum
RTD with a span of 100°F (50°F minimum) calibrated in our sensor-matching
calibration bath.
January 2012
Data Sheet 3.80
Page 1
Mounting choices include
dual-compartment and affordable
single compartment field
enclosures, compact connection
heads and a high-density
DIN-style housing.
Figure 1. Universal, programmable input with a fully isolated and
linear analog output.
Description
Moore Industries’ Smart HART
®
Temperature
Transmitters configure in minutes to accept a direct
signal input from nearly every available RTD and
thermocouple, as well as from ohm and mV sources.
Universal and microprocessor-based, they provide
an isolated and linear 4-20mA output proportional
to the input. This signal is ready for direct interface
with HART or non-HART based DCS, PLC and other
computer-based SCADA systems.
Easy-to-Order Temperature Assemblies
One simple model number is all it takes to order our
complete temperature assemblies. They come con-
figured, calibrated, wired and ready to install in your
process. See our RTI1 and RTI2 Ready-to-Install
Temperature Transmitter Assemblies data sheets
for details.
Features
• Input-to-output analog accuracy of up to
±0.014°C (±0.025°F)* is the absolute best in the
industry.
• 20-bit input resolution delivers exceptional
digital accuracy of ±0.1°C (±0.18°F) with all
Pt RTDs, or up to ±0.05°C (±0.09°F)* for Pt1000
RTD.
• Set up with HART Communicator, HART-based
system, or PC allows you to check the status, or
perform parameter changes, from the control room
or any field termination point on the 4-20mA wires.
• Long-term stability provides up to 5 years
between scheduled calibrations.
• Standard integral displays show real-time
process status and valuable loop diagnostic
information.
• AdvancedEMI/RFIprotectionandambient
temperature compensation guard against
environmental factors that can quickly degrade
measurement accuracy.
• PCsoftware’s“HelpMapNavigationSystem”
provides quick and complete answers to
performance, setup, installation and maintenance
questions.
Certifications
ADDR
0
100
Isolated,
Scaleable
4-20mA
with HART
®
Superimposed
Programmable
RTD
Thermocouple
Millivolt
Resistance
Potentiometer
2-Wire
(Loop-Powered)
235-710-11G
© 2012 Moore Industries-International, Inc.
All product names are registered trademarks of their respective companies.
HART is a registered trademark of the HART Communication Foundation.
Page view 0
1 2 3 4 5 6 ... 15 16

Summary of Contents

Page 1 - THZ & TDZ

THZ & TDZSmart HART® Temperature Transmitters*High-accuracy measurements are achieved by using a 4-wire, 1000 ohm platinum RTD with a span of 100

Page 2

THZ & TDZSmart HART® Temperature TransmittersPage 10Page 12Page 13Page 14Page 14Page 15BH Page 15D-BOXPage 16VersatileHousing,EnclosureandMoun

Page 3

THZ & TDZSmart HART® Temperature TransmittersPage 11Ordering InformationUnitTHZ Smart HART Tem -perature Trans-mitter With and Without Display

Page 4

THZ & TDZSmart HART® Temperature TransmittersPage 1249mm(1.93 in)ADDR0100GND1/2 NPT102mm(4.00 in)154mm(6.06 in)66mm(2.60 in)SIDE VIEW (NPT-PORT SI

Page 5

THZ & TDZSmart HART® Temperature TransmittersPage 1392mm(3.61 in)9mm(0.35 in)87mm(3.44 in)ConduitEntry Port89mm(3.50 in)84mm(3.30 in)CL61mm(2.40 i

Page 6

THZ & TDZSmart HART® Temperature TransmittersPage 1440mm–PS+PS1234TOP BOTTOMSIDECLCLCLCL(1.56 in)25mm(1.0 in)52mm(2.04 in)49mm(1.92 in)27mm(1.05 i

Page 7

THZ & TDZSmart HART® Temperature TransmittersPage 15FRONT VIEWSIDE VIEW61mm(2.40 in)+PS-PS 1234TDZ83mm(3.25 in)63mm(2.45 in)64mm(2.53 in)66mm(2.58

Page 8

THZ & TDZSmart HART® Temperature TransmittersPage 16Canadian Standards Association (CSA Int’l) Intrinsically Safe [TDZ-HP, THZ-HPP] – Class

Page 9

THZ & TDZSmart HART® Temperature TransmittersPage 2 Set Up with HART Communicator, DCS, AssetManagementSystem(AMS)orPC Our Smart HART Transm

Page 10

THZ & TDZSmart HART® Temperature TransmittersPage 3–+–+–+HART-Based DCS (Primary Master)HARTCommunicator(HARTSecondaryMaster)THZ(HARTSlave)THZ (HA

Page 11

THZ & TDZSmart HART® Temperature TransmittersPage 4TrimstoRespondtoSpecic Sensor Curve Segments Most transmitter’s zero and span values can

Page 12

THZ & TDZSmart HART® Temperature TransmittersPage 5IntelligentPCCongurationSoftware Our Intelligent PC Configuration software allows you (with

Page 13

THZ & TDZSmart HART® Temperature TransmittersPage 6GuidesYouFromHook-UptoStart-UpJust click the Help icon on the toolbar of our Intelligent P

Page 14

THZ & TDZSmart HART® Temperature TransmittersPage 7THZ-HPP T/C, mV THZ-HPP RTD, Ohm, Potentiometer THZ-DIN T/C, mV THZ-DIN RTD, Ohm, Potentiome

Page 15

THZ & TDZSmart HART® Temperature TransmittersPage 8Specications(continued)Display (THZ [DH] & TDZ only)THZ in DH Enclosure:Top Row, 18mm (0

Page 16

THZ & TDZSmart HART® Temperature TransmittersPage 9100 200 300 400 500 1000 100 200 400 500 1000 100 120 9.035 0-4000 ohms 4000

Comments to this Manuals

No comments